Share optimized solutions, professional valve knowledge and industry news

Please enter the relevant terms or keywords you need to consult, and relevant articles will appear in the search results. If you can’t find the answer you need, please feel free to contact us and we will be happy to help. Or you can directly send an email to [email protected]

Fisher Positioner FIELDVUE DVC62000 Auto Travel Calibration Using AMS Trex

Table of Contents

This quick guide explains how to perform an automatic travel calibration of a Fisher FIELDVUE DVC6200 digital valve controller using an Emerson AMS Trex communicator.

Before You Start

Make sure that:

  • The DVC6200 is correctly mounted on the actuator.
  • The travel feedback linkage or magnetic array is correctly installed.
  • Instrument air and the 4–20 mA signal are available.
  • The actuator and relay configuration are correct.
  • The valve can safely move through its full travel.

Warning: Auto calibration moves the valve through its full stroke. Isolate the valve from the process and release or equalize process pressure before starting.

wiring dvc6200
wiring dvc6200

DVC6200 Auto Calibration Procedure

StepProcedure
1. Connect the AMS TrexConnect the Trex HART leads to the DVC6200 signal terminals or to a suitable HART connection point in the 4–20 mA loop.
2. Connect to the PositionerSelect HART on the Trex and tap Connect. Select the DVC6200 and enter the Online menu.
3. Check the Valve and Actuator SetupGo to Configure → Manual Setup → Valve and Actuator. Confirm the actuator type, travel sensor motion, valve action and installed relay type.
4. Place the DVC6200 Out of ServiceGo to Configure → Manual Setup → Mode and Protection → Change Instrument Mode. Select Out of Service and confirm. Set instrument protection to None when required. The DVC6200 must be Out of Service and unprotected before communicator-based calibration.
5. Open Auto Travel CalibrationGo to Configure → Calibration → Travel Calibration → Auto Calibration. Menu wording may differ slightly depending on the HART and device-description version.
6. Start the CalibrationFollow the instructions displayed on the Trex. Confirm that the valve is safe to stroke, then start the calibration. The DVC6200 will move the valve through its travel range and determine the calibrated travel limits.
7. Complete Relay Adjustment, If RequestedFor a double-acting actuator using Relay A, the Trex may ask whether relay adjustment should be performed. Follow the on-screen instructions or select No to continue when the relay has already been correctly adjusted.
8. Confirm the Calibration ResultWait until the Trex reports that auto calibration has been completed successfully. Select OK or Done to exit the calibration procedure.
9. Return the Positioner to ServiceGo to Mode and Protection → Change Instrument Mode and select In Service. This allows the DVC6200 to resume normal control from the 4–20 mA input signal.
10. Verify Valve OperationApply several input signals, such as 0%, 25%, 50%, 75% and 100%. Confirm that the actual valve travel follows the commanded position and that no active calibration or travel alerts remain.

Relay Type Selection

The relay setting entered in the Trex must match the relay physically installed in the DVC6200.

Relay TypeApplication
Relay ADouble-acting actuators or single-acting direct applications
Relay BSingle-acting reverse applications
Relay CSingle-acting direct applications

For a single-acting direct configuration, actuator pressure is connected to Output A. For a single-acting reverse configuration using Relay B, actuator pressure is connected to Output B.

Important: Do not select Relay A, B or C only according to the required fail-open or fail-close action. Check the relay label, pneumatic output connection and actuator configuration before calibration.

Checking the Last Calibration Result

To review the previous automatic calibration result, go to:

Configure → Calibration → Travel Calibration → Last AutoCal Status

The calibration menu may also include:

  • Manual Calibration
  • Calibration Record
  • Calibration Type
  • Calibration Time
  • Calibrator

Available items depend on the DVC6200 HART version and device configuration.

Common Calibration Problems

Auto Calibration Does Not Start

Check whether:

  • Instrument Mode is set to Out of Service.
  • Instrument protection is set to None.
  • The DVC6200 has sufficient loop current.
  • HART communication is stable.
  • Supply air pressure is sufficient.
  • The valve is free to move.

Valve Moves in the Wrong Direction

Check:

  • Relay type.
  • Output A or Output B tubing connection.
  • Actuator type.
  • Travel sensor motion setting.
  • Zero-power condition.
  • Valve opening direction.

Calibration Fails Before Reaching Full Travel

Possible causes include:

  • Insufficient supply pressure.
  • Incorrect actuator size or spring range.
  • Mechanical obstruction.
  • Excessive valve friction.
  • Incorrect feedback installation.
  • Travel sensor moving outside its valid range.
  • Incorrect relay adjustment.

Quick Menu Path

Connect → Online → Configure → Manual Setup → Mode and Protection → Out of Service

Then:

Configure → Calibration → Travel Calibration → Auto Calibration

After calibration:

Mode and Protection → In Service

Recent Posts

Facebook
Twitter
LinkedIn
Pinterest
Print
Picture of Will Don

Will Don

After earning my bachelor's degree in mechanical engineering from Zhejiang Normal University in 2008, l was fortunate enough to begin my career with Siemens, Fisher, and YTC, focusing on control valve accessories. Over the past dozen years, l've poured my heart and energy into understanding technology and fluid solutions for control valves.
Now, as the marketing director for THINKTANK, a trusted branch of the Taiwan STONE valve group, I can't help but feel proud of how far we've come.
Our knowledge isn't just reaching professionals like engineer and valve distributors; it's also inspiring the next generation of automation college students.
l genuinely hope you're enjoying our articles and finding them helpful. Your thoughts, questions, and feedback mean the world to me, so please don't hesitate to reach out to [email protected]. Whether you're a seasoned expert or just curious about the field, I'm here to connect, share, and learn together.

All Posts »

I am the author of this article, and also the CEO and marketing director of THINKTANK, with 15 years of experience in the industrial valve industry. If you have any questions, you can contact me at any time.

Contact Us

Just fill out your name, email address, and a brief description of your inquiry in this form. We will contact you within 24 hours.

Why are more and more international valve brands choosing THINKTANK for OEM service?

Brand is reputation, which not only requires high valve technology, but also great comercial support and fast response after-sales service and delivery time. They will not gamble their brand on an unknown product

Get an Instant Quote

Avoid your inquiry is delay response, please enter your WhatsApp/Wechat/Skype along with the message, so we can contact you at the very first time.

We will reply you within 24 hours. If for urgent case, please add WhatsApp: +86 185 1656 9221, or WeChat: +86 199 2125 0077. or call +86 189 5813 8289 directly. 

Get An Instant Quote

Joy Preview

Get An Instant Quote

We will reply you within 24 hours. If for urgent case, please add WhatsApp: +86 199 2125 0077, or WeChat: +86 199 2125 0077. Or call +86 189 5813 8289 directly. 

Direct to Control Valve Factory With Competitive Price & Reliable Quality

Joy Preview
Joy Preview

Get An Instant Quote

Just leave your name, email, and simple message or requirements, We will contact you within 1 hour.

WhatsApp: +86 199 2125 0077

Skype ID: sowell85

Wechat ID: +86 199 2125 0077